Difference between revisions of "CITI-Sense-Mob sensor testing"

From its-wiki.no
Jump to: navigation, search
(Status)
(Status)
Line 43: Line 43:
 
* static electricity discharging.
 
* static electricity discharging.
 
* Partial testing for different parts of the sensor.
 
* Partial testing for different parts of the sensor.
 +
  
 
caution:
 
caution:
sensor used techs for air contents measurement based on the using of ACID (chemical substance) which sometimes cause bad smell.
+
sensor used techs for air contents measurement based on the using of ACID (harmful chemical substance) which sometimes cause bad smell.
  
 
Needed for future testing:
 
Needed for future testing:
 
* list of used commands for sensor configuration in order to simplify and accelerate testing.
 
* list of used commands for sensor configuration in order to simplify and accelerate testing.
 
* Mobile SIM card for sensors configuration.
 
* Mobile SIM card for sensors configuration.
 +
* Sensor manual
  
 
= Dunavnet link =
 
= Dunavnet link =

Revision as of 08:01, 6 March 2014


CITI-Sense-Mob sensor testing

Title CITI-Sense-Mob sensor testing
Place UNIK
Date, Time 2014/03/03, 0900-1100
Contact Person Josef Noll
Participants Seraj Fayyad
related to Project Citi-Sense-MOB
Keywords
this page was created by Special:FormEdit/Meeting, and can be edited by Special:FormEdit/Meeting/CITI-Sense-Mob sensor testing
Category:Meeting



Status

1) earlier tests failed, no sensor signal

Potential reasons

  • no power supply, - measure with multimeter
  • to little power (effect) to fire up the sensor board - - don't know the power
  • communication fails

On-device monitoring

  • Tested the sensor with usb directly to pc and checked that there is some communication in a terminal program
  • Also, open it and recheck if the sim card is correctly in the slot.

2) second test partially success, no location information, show some measured values

Potential reasons:

  • static electricity

Done activities:

  • Send SMS to sensor SIM card (40923931), content =<SRVADR>srv.dunavnet.eu</SRVADR>
  (goal): change server address to the new one.
  • Sensor discharging from static electricity.
  • Sensor running without SIM card then with SIM card in order
  (goal): force sensor to recheck the SIM card existence

3) third test: Temporal success with location identification and measurements :

location identification failed potential reasons:

  • some wire bad connectivity.

Done activities:

  • recheck wires connectivity
  • static electricity discharging.
  • Partial testing for different parts of the sensor.


caution: sensor used techs for air contents measurement based on the using of ACID (harmful chemical substance) which sometimes cause bad smell.

Needed for future testing:

  • list of used commands for sensor configuration in order to simplify and accelerate testing.
  • Mobile SIM card for sensors configuration.
  • Sensor manual

Dunavnet link

Dunavnet 3 online data

Screenshot Dunavnet Sensor3.png

GUI


Last measured:


Link status

  • lab testing 20Dec2013, position: 59.9755117, 11.0535750 (NILU lab)

Position external sensor3.png


internal sensor measures: Air_Quality_bike_sensor_parameter